Detachable Implant 20: 1 Contra Angle Handpiece
Shanghai Woods Packaging Co., Ltd. / 2022-06-09
- Condition:New
- Certification:TUV, ISO, CE
- Material:Titanium
- Light Way:Without Light
- Water Way:Outer Channel
- Usage Times:Non-Disposable
=== Base Info ===
- Model NO.:R20N-01C
- Application:Adult
- Loading Way:Gland Type
- Head Diameter:Middle Head
- Exhaust Way:Exhaust Outside
- Number of Holes:Two Holes
- Connection:Without Quick Connector
- Rotating Speed:Low-Speed Handpiece
- Gear Ratio:20:1 Reduction
- Noise:Less Than 70dB
- Transport Package:Plastic Box
- Specification:Aluminum, stainless steel
- Trademark:TEALTH
- Origin:China
- HS Code:90184100
- Production Capacity:50000PCS,Year
=== Description ===
Basic Info.
Model NO.
R20N-01C
Application
Adult
Loading Way
Gland Type
Head Diameter
Middle Head
Exhaust Way
Exhaust Outside
Number of Holes
Two Holes
Connection
Without Quick Connector
Rotating Speed
Low-Speed Handpiece
Gear Ratio
20:1 Reduction
Noise
Less Than 70dB
Transport Package
Plastic Box
Specification
Aluminum, stainless steel
Trademark
TEALTH
Origin
China
HS Code
90184100
Production Capacity
50000PCS/Year
Product Description
Name: NSK SG20 20:1 implant contra angle dental handpieceOrder code: R20N-01C
Pack list :
1.20:1 contra angle *1 pcs
2.Cap spanner * 1 pcs
3.Operation manual * 1 pcs
4.E -type Oil spray nozzle *1 pcs
5.Internal water tube * 1 pcs
6.Tube fixture holder *1pcs
Parameter:
Titanium body appearance
Push button type
torque force: up to 55 N.cm
Rotation speed :5~2000RPM
ergonomics coating ,more hygeian.
Sterilization : up to 135 ºC
Applicable bur : Φ2.345-2.355mm
Fitting with E type motor , can be avaible to match up with Swiss, Japan, Korea famous brand implant system.
Insert or remove bur or drill
1) Inserting(Fig.1,Fig2)
Press the push button firmly, insert bur and turn the rod in clockwise direction, till the bur to the end.
Removal
Press the push button firmly and remove the bur.(Fig.3)
Handpiece inserting & removal:
Inserting
Insert the handpiece into the motor and turning all the way until it click to stop.(Fig.4)
2) Removal
Grip both handpiece and motor respectively,pulling and separating.
Notes:
Please do not insert or remove handpiece before motor complete stop turning.
Compatible with dental motor implant system in accordance with ISO standard.
Compatible with the electro-motor .
The structure of connector conform with ISO3964 standard.
Water Spray:
1) External water spray
Connect the water pipe and the external water spray pipe of handpiece.(Fig.5)
2) Internal water spray
Use internal bur to operate. Water pipe connect to the internal water spray pipe, and make the internal water spray pipe fixed with the pipe fixture. (Fig.6)
3) Mixed water spray
Use internal bur to operate, and follow step 1 and 2.
Maintenance & Care:
1. Cleaning
Remove bur from handpiece after using and brush. soak the head into clean water and operate for 15 seconds(Fig.7). Then autoclave handpiece.
If the handpiece is the type of head can remove one, please take off the head and wash it and the body with clean water.
2. lubricating(Fig.8)
Lubricate oil with E type oil nipple insert the handpiece and spray 2 or 3 seconds.
If handpiece head dirty enough, please remove the back cap with spanner and spray oil inside(Fig.9 and 10).
Please lubricate handpiece per 30 minutes while working continuous.
The handpiece shall be completely cleaned after several times use. Twist the handpiece and take off the upper part, wrench the bearing with bearing spanner and clearing the cartridge to keep the performance of the contra angle(Fig.11, the model of head can remove do not need this step).
3. Sterilization
It shall be sterilized before handpiece is used at the first time: Autoclave the handpiece for 12 min. at 135ºC
After each treatment, sterilizing handpiece in autoclave according to the following steps:
1)Remove the bur and brush clean handpiece, clean outside of handpiece with an alcohol solution.
2)Lubricate handpiece.
3)Place into autoclaving pouch and mark.
4)Autoclave the handpiece for 12 min. at 135ºC.
Note:
Please do not use the acid solution to scrub, inject or soak the handpiece.
Please do not autoclave the handpiece with residual chemical substances.
Handpiece should be packed well and place at clean and dry room without dust, sulfur and salt.
Please do not touch the handpiece immediately after autoclaving.